Operating conditions for gauge pressure and absolute pressure with
flush-mounted diaphragm
Ambient conditions
Ambient temperature
Note Observe the temperature class in hazardous areas.
Enclosure
Measuring cell with sili‐
cone oil filling
-40 … +85 °C (-40 … +185 °F)
Measuring cell with inert
oil (different pressure
classes)
1 bar/100 kPa/3.6 psi -40 … +100 °C (-40 … +212 °F)
4 bar/400 kPa/58 psi -40 … +100 °C (-40 … +212 °F)
16 bar/1.6 MPa/232 psi -40 … +100 °C (-40 … +212 °F)
63 bar/6.3 MPa/914 psi -40 … +100 °C (-40 … +212 °F)
160 bar/16 MPa/2321 psi -20 … +100 °C (-4 … +212 °F)
400 bar/40 MPa/5802 psi -20 … +100 °C (-4 … +212 °F)
700 bar/70 MPa/10152 psi -20 … +100 °C (-4 … +212 °F)
Measuring cell with FDA-
compliant oil
-10 … +85 °C (14 … 185 °F)
Display -20 … +80 °C (-4 … +176 °F)
Storage temperature
-50 … +85 °C (-58 … +185 °F)
(with FDA-compliant oil: -20 ... + 85
°C (-4 ... +185 °F))
Climate class in accordance
with IEC 60721-3-4
4K4H
Degree of protection accord‐
ing to IEC/EN 60529/UL50-E
Enclosure with appropriate cable
gland
IP66/Type 4X
IP68 (2 hours at 1.5 m)
Enclosure with mounted M12 con‐
nector and/or lightning protection
IP66/Type 4X
Enclosure with mounted HAN con‐
nector
IP65
Electromagnetic compatibili‐
ty
Interference emission and
interference immunity
In accordance with EN 61326 and
NAMUR NE 21
Process medium conditions
Process temperature
1)
Measuring cell with
silicone oil filling
-40 … +150°C (-40 … +302 °F)
-40 … +200°C (-40 … +392 °F) with
cooling extension
Measuring cell with inert oil
-20 … +100 °C (-4 … +212 °F)
Measuring cell with FDA-
compliant oil
-10 … +150°C (14 … 302 °F)
-10 … +200°C (14 … 392 °F) with
cooling extension
1)
Observe the temperature limits in the process connection standards (e.g. DIN 32676 and DIN 11851) for the maximum
process temperature for flush-mounted process connections.