Operation Function Tests
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
D.2-3
0.0
Checking the table top
D. 2
You must check the mobility and cleanness of the table top.
D. 2
Make sure that you can manually retract the table top from
the gantry. (→ Page C.2-9 Retracting the patient table
manually)
Make sure that the scan field is free of residual contrast
medium, blood or other contaminations.
Checking the laser light marker
D. 2
D. 2
Switch on the laser light marker and check the projections
of the light beams on a white sheet of paper.
The laser beam must project a line.
D. 2
fB
D.2
The laser light marker does not work.
D. 2
Stop scanning to rule out any danger to patients.
Call Siemens Customer Service.
D. 2