syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Function Tests Operation
D.2-4
Checking the intercom system
D. 2
Apply the Hear Pat (Hear Patient) and Call Pat (Call Patient)
keys on the control box. (→ Page B.5-11 Control box)
The intercom system must work in both directions.
D. 2
Checking radiation block and
warning lamps
D. 2
You must check the radiation warning lamps on the control box,
on the gantry and, if present, next to the doors of the examina-
tion room.
D. 2
Check the proper function of the radiation warning lamps
during the daily image quality tests.
The radiation warning lamps must light up when radiation is
generated.
D. 2
Make sure that radiation stops immediately when a door of
the examination room is opened
i
D.2
When the door is closed again you must restore system readi-
ness with System > Continue.
D. 2