Operation Function Tests
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
D.2-5
0.0
Checking the STOP keys
D. 2
D. 2
Press one of the table setting keys on the gantry operator
panel and hold it down.
D. 2
When the table moves, press a STOP key.
The table movement must stop immediately.
D. 2
Check that you can move the table out of the gantry man-
ually after you have pressed a STOP key.
i
D.2
After you have applied a STOP key, you must restore system
readiness with System > Continue.
D. 2