Quality Assurance General Information
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.1-5
0.0
i
.1
F
The numerical value for the correct table height will be dis-
played after the phantom position check.
F.1
F.1
Move the patient table into the gantry.
The beam of the light marker must be aligned with the refer-
ence marking of the slice phantom.
F.1