Quality Assurance Daily Quality Measurements
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.2-5
0.0
Triggering radiation
F.2
In the Quality Daily dialog box you are prompted to initiate
radiation.
F.2
F.2
Press the Start key on the control box.
The position of the phantom set is checked and displayed. Then,
the first measurement is started automatically.
F.2
i
.2
F
For incorrect positioning of the phantom set, a message is dis-
played. Correct the position, if necessary.
F.2
For each tube voltage (kV step), two measurements are per-
formed.
F.2