syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Daily Quality Measurements Quality Assurance
F.2-4
Activating daily quality test
F.2
✧ Call up Setup > Quality.
The Quality Daily dialog box is displayed.
F.2
F.2
Starting measurement
F.2
F.2
✧ Click on the OK button.
The tilt of the gantry is automatically set to 0°, if necessary.
F.2
The “Quality Assurance Patient” is automatically selected.
F.2
The patient table is positioned such that the water phantom is
in the scan plane.
F.2