Constancy Test Quality Assurance
F.3-26
Performing the table posi-
tion test
F.3
With this quality measurement, you test whether the actual
table position matches the table position displayed.
F.3
The Table Position test is performed without triggering radia-
tion. It is started with Go in the Quality Constancy dialog box.
(→ Page F.3-2 Performing the constancy test)
F.3
i
.3
F
After the previous test has been completed, the Table Position
test is started automatically.
F.3
✧ Mount a ruler at the mobile part of the patient table so that
the 0 mm mark of the ruler is aligned with the stationary
part of the table.
✧ Place the 135 kg weight on the patient table.
F.3
✧ Start measurement with Go.
The horizontal table position is set to 0. The table top moves
300 mm in the direction of the gantry.