Quality Assurance Constancy Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.3-27
0.0
✧ Read the 300 mm position on the ruler.
F.3
✧ Enter the value in the entry field of the content area.
F.3
✧ Click on Continue.
The table top now moves stepwise into the gantry (300 mm)
and back.
F.3
✧ Enter the actual values at the 300 mm and the 0 mm posi-
tion in the entry field again.