Quality Assurance CTDI Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.4-5
0.0
Starting the test
F.4
F.4
In the Quality Constancy dialog box, select the individual
tests CTDI Head (opt.) and CTDI Body (opt.).
(→ Page F.3-2 Performing the constancy test)
F.4
Confirm selection of the constancy test with Go.
You are prompted to enter the name of the tester and the serial
numbers of the phantoms.
F.4
F.4
Enter your name and the data required.
F.4
Start the CTDI test with Go.
i
.4
F
If you are performing the CDTI measurements as part of the
monthly constancy test, replace the set of phantoms with the
CTDI head phantom used for the constancy test now.
F.4
F.4
Confirm the correct phantom position with Go.
You are prompted to press Start.
F.4