syngo CT 2007E Emotion Operator Manual
02/2008
0.0
CTDI Test Quality Assurance
F.4-6
F.4
Press the Start key on the control box.
The phantom position is checked. The tomographic image of
the phantom is displayed in the topo segment.
F.4
F.4
In this example, the head phantom is shown.
F.4
Check whether the center hole of the phantom lies within
the drawn in ROI.
i
.4
F
If the phantom is not located in the scan center, correct its posi-
tion. Check the phantom position again with Repeat.
F.4