Quality Assurance Low Contrast Test (Optional)
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.5-3
0.0
Starting the test
F.5
The Low Contrast test is performed in the Quality Constancy dia-
log box of the Local Service window.
F.5
✧ Call up the Quality Constancy dialog box.
F.5
✧ Select the Phantom Check and the Low Contrast test.
F.5
✧ Confirm selection of the constancy test with Go.
You are prompted to enter the name of the tester and the serial
numbers of the phantoms.
F.5
✧ Enter your name and the data required.