Sirona Dental Systems GmbH 10 Operating light, tumbler filling, flushing of the cuspidor, purge
C3
+
10.4 Purge / Auto Purge function (purging the water paths)
59 06 404 D 3390
D 3390.201.01.18.02
63
Setting the APURGE function
Activate the selection dialog box by pressing quick
adjustment key 1 (> 3 s) when all of the instruments are
deposited in their holders.
You will be prompted to choose between PURGE/
APURGE and SANITATION
1
.
Select APURGE with the
– or + key. APURGE must
flash.
When you press quick adjustment key 1 briefly (< 3 s),
the following prompt appears if all instruments are in
their holders:
SELECT ITEMS FOR PURGING