10 Operating light, tumbler filling, flushing of the cuspidor, purge Sirona Dental Systems GmbH
10.4 Purge / Auto Purge function (purging the water paths) C3
+
59 06 404 D 3390
64 D 3390.201.01.18.02
Remove the instruments to be purged from the dentist
and assistant elements and insert them into the adapters
of the sanitation bowl as far as they will go.
Be careful not to snap off the instrument hoses when
inserting the instruments.
The maximum water flow must be set on the instru-
ments.
Press quick adjustment key 1 again (< 3 s); the instru-
ments are then purged one after the other.
The purging time in seconds set in section "Mode:
PURGE TIME... SEC" on page 71 runs down for each
instrument individually.
An audible signal sounds after the last instrument has
been purged.
You now can continue in either of the following ways:
When removing the sanitation bowl, please note that it is
filled with water.
•
Leave the instruments in the sanitation bowl:
The treatment center can then be switched off.
The instruments will automatically be purged as
soon as the treatment center is switched on
again.
The APURGE function remains activated.
•
The instruments are returned to their holders:
The treatment center can be prepared for daily
practice operation.
The APURGE function is completed.