SKOPE BME-N ActiveCore
Troubleshooting
Service Manual
• Product is too warm • Frequent door opening. • Limit door openings.
• Recently loaded. • Allow time for the product to cool down.
• Door not closing
properly.
• Check and the clean door gasket.
• Refrigeration cartridge
operating too hot.
• Excessive door opening
or refrigeration heat load.
• Ensure the cabinet has good ventilation
around the refrigeration cartridge.
• Ensure the cabinet is within the maximum
operating conditions.
• Set point is too high. • Lower the set point.
Moisture build up on door
or exterior
• High humidity. • Check ambient operating temperature and
reposition cabinet if necessary.
• Frequent door opening. • Limit door openings.
• Door not closing
properly.
• Check and clean door gasket.
• Cabinet door does not
shut properly.
• Cabinet is on an uneven
surface.
• Level the cabinet.
• Door is obstructed. • Check shelves and product.
• Warm cabinet
temperatures
• Compressor operating
for long periods (more
than 1 hour)
• Blocked condenser • Clean the condenser.
• Poor ventilation around
refrigeration cartridge
• Ensure the cabinet has good ventilation
around the refrigeration cartridge.
• Ensure the cabinet is within the maximum
operating temperature.
Table 18: Troubleshooting (continued)
Problem Possible cause Repair