2-3
2.2 Specifications and Installation Characteristics
Dimensions: Height:
Width:
Depth:
300 mm
223 mm
285mm
Weight: Analyzer:
5.5 kg (12.1 pounds)
Ambient Operating Temperature:
Operating Humidity:
Device Storage/Transportation
Temperature:
Device Storage/Transportation
Humidity:
Reaction Temperature:
Thermal Protection Rating:
Waterproof Class:
Altitude:
Power Requirements:
Main Supply Voltage:
Transient Over voltages:
Pollution:
Consumables:
Quality Control:
Samples:
Sample Volume:
Warm-up Time:
Test Time:
Touch Panel:
Printer:
Communication:
Memory Capacity:
Indoor 10 - 32C (50 - 89.6F)
<95% relative humidity (non-condensing)
<65C(149 F)
<90% relative humidity (non-condensing)
37C (98.6 F)
100C (212 F)
Ordinary equipment (IPX0)
2,000 m (6,562 ft)
100-240 volts AC; 50-60 Hz; Output 12 Volts DC; 5.0 A
Fluctuations must not exceed ± 10% of the nominal voltage
Installation Category II in accordance with EN610101-1
Degree 2 in accordance with IEC 664
Reagent disc/pipette tip
Self-calibration function/internal quality control function for each reagent disc
Serum and lithium-heparinized whole blood or plasma
200 µL ±10 μL
Less than 15 minutes (may exceed this time limit in unique environments)
15 minutes
5” TFT color LCD touchscreen
Built-in thermal printer; can be connected to compatible external printers
USB port: 1x ; RS232 port: 1x ; RJ45 network port: 1x
Capacity for up to 50,000 test results