3-20
Notes:
1. Perform and record the
quality control test
according to the
laboratory’s quality control
regulations. Quality control
records should be
indefinitely documented in
the laboratory’s records.
2. The quality control tests
must be performed using
the Control function in the
Settings on the skyla
TM
Clinical Chemistry
Analyzer to ensure the
accuracy of the results.
3. Please refer to the package
insert of the selected
control for detailed sample
preparation and testing
information.
Press SETTINGS on the main menu screen.
Press the Control icon to start the quality control
test
process. The test process is the same as the one for
sample analysis and
the system will automatically
complete the test.
However, there is no need to input
any patient related information.
The
quality control test results are stored separately
from sample analysis test results. T
he results can be
printed out or exported after the quality control test is
completed.
Quality control test results are stored in the system
database. You can search for, review, and print
past
control test results using the RECALL
function (see
Section 4.1.3
Searching for System Quality Control
Test Records’.
3.9.4 Quality Control Test Results
Quality control test results of each test marker should
be
in agreement with the reference values provided by the
control manufacturer.