3-15
Symbols used in test reports:
SCORE: Indicates the
numerical value of sample
lipemia (LIP), hemolysis
(HEM), and icterus (ICT).
minIndicates the
minimum acceptable value
Table 2: Test report generated when the QC Report
function is activated
3.7.2 Symbols Used in internal QC Reports
QC1: This QC item indicates whether the volume of
the diluted sample is sufficient. Normal score
range is
90 or over.
QC2: This QC item indicates whether the diluent has
been contaminated. Normal score range is 90 or over.
QC3: This QC item indicates whether the assay
chemical reagents have been degraded
due to
deliquescence. Normal score range is 90 or over.
QC4: This QC
item indicates whether the dilution is
sufficient. Normal score range is 90 or over.
QC5: This parameter indicates whether the volume of the
sample is sufficient. Normal score range is 90 or over.

Table of Contents