2-11
Power-Off:
Press the power icon on
screen and pressYES to
confirm power-off.
Prior to inserting a reagent disc into
the clinical
chemistry analyzer
, check that the expected reference
ranges and specified units have been set up properly.
Also check if there is thermal paper in the built-
in
thermal printer, or, if it applies, whether setup of
the
external printer and other connected devices are
complete.
Power-Off
Please turn off the clinical chemistry analyzer
if it will
not be used for a prolonged period. Press the power icon
on screen, and press “YES” to confirm power-off.
2.6 Maintenance and cleaning
LITE-ON suggests the skyla
TM
Clinical
Chemistry Analyzer
should be cleaned routinely
at least once per week.
Use a clean cloth dampened with cleaning
solution (ex. 10% commercial bleach solution) to
wipe the
exterior surface of the analyzer. When
using the cleaning solution for the first time
gently wipe a small area of the analyzer surface
to ensure no obvious damage occurs before
cleaning the entire surface.
If the exterior surface of analyzer is stained wit
h
blood then clean immediately using a cloth
dampened with 10% commercial bleach solution.
When necessary, clean the interior area of disc
drawer using the following steps:
1.
Press the “STARTicon on screen to open the
disc drawer.
2. Use a cotton swab d
ampened with 10% commercial
bleach solution to wipe the drawer area.