6-1
Troubleshooting:
■ Power and Function
■ Settings and Reports
■ Data Export
■ Password Setting
Section 6: Troubleshooting & Error
Handling
Operation of the skyla
TM
Clinical Chemistry Analyzer
simple
and almost fully automated. When a user error
occurs during operation, the system automatically
displays an error message. Most issues which users will
encounter can be easily resolved by following the simple
steps in the system’s suggestions.
6.1 Troubleshooting
The troubleshooting recommendations contained in this
section are solutions (S) to resolve common conditions
(C) users encounter in a few simple steps.
■ Analyzer Power and Function
C: Touchscreen is black or blue; power light is OFF.
S: Turn the analyzer power off.
Check the power
cable connections. Disconnect and reconnect on
both ends, then turn the power on.
C: Touchscreen is blue or blank; power light is ON.
S:
Remove the reagent disc from the drawer. Check
that the drawer closes properly, then re-
analyzer.
C: Damaged touchscreen
S:
Please contact the authorized distributor or
technical support department of
LITE-
Technology Corp.
.
■ Settings and Reports
C: There are no values for some
of the results within a
printed test report, or the system automatically
cancels a reagent disc test without generating
report.