6-2
S: The skyla
TM
Clinical Chemistry Analyzer per
forms
automatic system calibrations to maintain high
system precision. Thus, when the Analyzer detects
a faulty reagent disc, it automatically cancels the
test and displays an operational error message.
Please insert a new reagent disc to resume
operations.
C: The test report exported by the built-
in thermal
printer is blank.
S:
The thermal paper roll may have been installed in
the wrong direction within the built-
in thermal
printer. Please remove the roll of thermal paper and
confirm that the heat-sensiti
ve side is facing the
correct direction (the arrow sticker on the roll
should point downwards) before replacing the roll
in the built-in thermal printer.
C: After choosing to initiate the “System Default
function, previously customized reference range
settings have disappeared.
S:
Once the System Default function is initiated,
the system will delete all custom settings and
reference ranges. Once the command is executed,
all settings will be reset to the default
system
settings. The settings that have
been deleted cannot
be recovered. However, none of the patient
test
results and calibration records will be deleted.
Warning: Once the System Default
function is
initiated, all custom settings and reference
ranges will be irreversibly deleted.
Exporting data from the system to other devices
C: Peripheral and media of the exported data.
S: Dump to USB
function: will export selected
historical data to an external USB drive.