QC6: This QC item indicates the degree of sample
hemolysis. Normal score range is 90 or over.
QC7: This QC item indicates temperature control
stability of the instrument. Normal score is within the
range 90110.
L340nm: This QC item indicates the linear stability of
the equipment’s first optical channel. Normal s
core is
within the range 90–110.
L405nm: This QC item indicates the linear stability of
the instrument’s second optical channel. Normal score
is within the range 90–110.
L450nm: This QC item indicates the linear stability of
the instruments third optical
channel. Normal score is
within the range 90–110.
L510nm: This QC item indicates the linear stability of
the instruments fourth optical channel. Normal score
is within the range 90–110.
L600nm: This QC item indicates the linear stability of
the instrument
s fifth optical channel. Normal score is
within the range 90–110.
L650nm: This QC item indicates the linear stability of
the instrument’s sixth optical channel. Normal score is
within the range 90–110.
L540nm: This QC item indicates the linear stability
of
the instruments seventh optical channel. Normal score
is within the range 90–110.
L940nm: This QC item indicates linear stability of the
instruments eighth optical channel. Normal score is
within the range 90–110.
System QC: This QC item indicates the
stability and
reliability of the instruments system.
Normal score
range is 90 or over.
System QC Acceptable Minimum: 90