3-5
Release the fingers that
grip the regent disc,
unwrap and discard the
tissue paper.
Inspect and do not use any
damaged reagent discs.
Once the pouch is opened,
use the disc within 20
minutes.
Before dispensing the
sample, remove the
aluminum strip covering
the sample port on the disc.
After removal, use the other hand to hold the reagent
disc by the edge.
Release the fingers that grip the regent disc,
and discard the tissue paper.
Inspect the reagent disc
abnormality.
Do Not Use Any Damaged Reagent
Discs.
After opening the pouch, the disc
within 20 minutes. Do not place the disc back
refrigerator for later use.
The surface of the reagent disc should be kept
clean at all times. Hold the disc by the edge
avoid contact with the reagent disc surface
optical measurement takes place.
Remove The Aluminum Strip
Before dispensing the sample, remove
strip of the diluent container
to ensure normal operation
of the clinical chemistry analyzer. Please follow these
steps to remove the aluminum strip:
Hold the reagent disc by
the edge and avoid touching
the barcode ring. Meanwhile, avoid contacting
top and bottom sides of the reac
located on the outer part of the reagent disc
holes next to the barcode). Remove
strip
of the diluent container, which covers the
sample port on the reagent disc, by gently pulling
end of the aluminum strip in an outward direction
along the surface of the reagent disc. Please
completely remove the aluminum strip
reagent disc.
Discard the removed aluminum strip.