(I): Icterus interference
effect is outside the
permissible range by 10%
or more. “_:_:I” is shown.
(_:_:I): Test marker affected
by interference due to
icterus (ICT).
(_:H:_): Test marker
affected by interference due
to hemolysis (HEM).
(L:_:_): Test marker
affected by interference due
to lipemia (LIP).
(L:H:I): Test marker
affected by interference due
to lipemia (LIP), hemolysis
(HEM), and icterus (ICT).
mg/dL” will be shown.
(+): Indicates the degree of
(HEM), and icterus (ICT) of the sample.
plus symbols (+) are shown, the more severe
interference found in the sample.
The degree of lipemia (LIP), hemolysis (HEM), i
(ICT) of the sample is classified as follows:
“0” (Clear or without interference)
“+” (Mild)
“++” (Moderate)
“+++” (Severe)
(L): Indicates that the marker
interference due to sample lipemia (LIP) that
the permissible range by 10% or more.
“L:_:_” is shown.
(H): Indicates that the marker
interference due to sample hemolysis (HEM) that
outside the permissible range by 10% or more.
“_:H:_” is shown.
(I): Indicates that the marker
interference due to sample icterus (ICT) that is ou
the permissible range by 10% or more.
“_:_:I” is shown.
(_:_:I): Indicates that the marker
interference due to sample icterus (ICT).
(_:H:_): Indicates that the marker
interference due to sample hemolysis (HEM).
(L:_:_): Indicates that the marker has b
interference due to sample lipemia (LIP).
(L:H:I): Indicates that the marker
has been affected by
interference due to sample lipemia (LIP), hemolysis
(HEM), and icterus (ICT).