EasyManua.ls Logo

Skyla HB1 - Page 33

Skyla HB1
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-13
(I): Icterus interference
effect is outside the
permissible range by 10%
or more. “_:_:I” is shown.
(_:_:I): Test marker affected
by interference due to
icterus (ICT).
(_:H:_): Test marker
affected by interference due
to hemolysis (HEM).
(L:_:_): Test marker
affected by interference due
to lipemia (LIP).
(L:H:I): Test marker
affected by interference due
to lipemia (LIP), hemolysis
(HEM), and icterus (ICT).
mg/dL will be shown.
(+): Indicates the degree of
lipemia (LIP), hemolysis
(HEM), and icterus (ICT) of the sample.
The more
plus symbols (+) are shown, the more severe
this
interference found in the sample.
The degree of lipemia (LIP), hemolysis (HEM), i
cterus
(ICT) of the sample is classified as follows:
“0 (Clear or without interference)
“+ (Mild)
“++ (Moderate)
“+++” (Severe)
(L): Indicates that the marker
has been affected by
interference due to sample lipemia (LIP) that
is outside
the permissible range by 10% or more.
“L:_:_is shown.
(H): Indicates that the marker
has been affected by
interference due to sample hemolysis (HEM) that
is
outside the permissible range by 10% or more.
“_:H:_is shown.
(I): Indicates that the marker
has been affected by
interference due to sample icterus (ICT) that is ou
tside
the permissible range by 10% or more.
“_:_:Iis shown.
(_:_:I): Indicates that the marker
has been affected by
interference due to sample icterus (ICT).
(_:H:_): Indicates that the marker
has been affected by
interference due to sample hemolysis (HEM).
(L:_:_): Indicates that the marker has b
een affected by
interference due to sample lipemia (LIP).
(L:H:I): Indicates that the marker
has been affected by
interference due to sample lipemia (LIP), hemolysis
(HEM), and icterus (ICT).

Table of Contents