B-3
38 101 Wavelength 650nm value.
39 90-110 Wavelength 650nm standard range.
40 98 Wavelength 940nm value.
41 90-110 Wavelength 940nm standard range.
42 100 System quality control (System QC) value.
43 min.90
System quality control (System QC) minimum acceptable
value.
44 123 Chemistry quality control (Chemistry QC) value.
45 min.90
Chemistry quality control (Chemistry QC) minimum
acceptable value.
46 0
LIP indication: LIP indications are 0, +, ++ or +++. 0
represents no interference, +, ++, +++ represents degrees
of seriousness.
47 ++
HEM indication: HEM indications are 0, +, ++ or +++. 0
represents no interference, +, ++, +++ represents degrees
of seriousness.
48 +
ICT indication: ICT indications are 0, +, ++ or +++. 0
represents no interference, +, ++, +++ represents degrees
of seriousness.
49 50 LIP indication value.
50 25 HEM indication value.
51 30 ICT indication value.
52 ALB
Marker name: If the first character is “#”, this indicates
that this marker is calculated from other markers.
53 5 Analysis result. If this cannot be calculated, “N.A.” will
be exported.
54 4.5 Lower limit of male normal reference range.
55 7.5 Upper limit of male normal reference range.
56 4.5 Lower limit of female normal reference range.
57 7.5 Upper limit of female normal reference range.
58 g/dL Analysis result unit.
59 1 0: Close custom field export. 1: Open custom field
export.