<3> If the thin cotton swabs moistened with alcohol cannot eliminate the
contaminant in the V-groove, the operator can use the tip of the optical fiber with
the coating layer being stripped to remove the dirty from the V-groove, then
repeat steps <2> .
5.2 Cleaning fiber clamp
<1> If the fiber clamp is contaminated by dust, the fiber gripping may have
problem. This will likely result in poor quality of the splice point. In routine
working process, the fiber clamp should be checked and cleanedregularly.
<2> Cleaning fiber clamp according to following steps: Open the wind cover.
Clean with a thin cotton swab moistened with alcohol the surface of fiber clamp,
then wipe with a dry cotton swab the clamp.
● Be careful not to touch the electrode tip.
● Do not use excessive force when cleaning V-groove,in case
damaging V-groove wall.