120
Shenzhen Skyshl Technology Co.,LTD.
Sales email: sales1@skyshl.net After Sales email:af@skyshl.net Website:www.skyshl.net
SS413F Optical Fiber Fusion Splicer Users Manual
V
state. If changing the time setting, refer to [Power Save] (Section 3.7.5).
Change the cleave angle, splice loss, offset angle limit
See [splice mode] operation (Section 3.2)
Error messages can be ignored.
Enter [ignored display] in [splice options] menu (Section 3.5) and refer to the
error messages allowed to be ignored.
Arc intensity and time can not be changed.
In Auto mode or Calibrate mode,these parameters can not be changed.
In the modes above,after executing the arc calibration, the splicer can get
sufficient arc intensity.
In Normal mode,arc intensity and time can be changed unless the operator
locks them.
Showing cleave angle, fiber angle, fiber core/ clad offset
For cleave angle,fiber angle, core / cladding offset settings, see [data display]
in [splice options] (Section 3.5).
In the automatic mode, unsuitable splice mode is selected.
Automatic mode can only apply to standard SM, DS, NZ, MM optical
fibers.See [splice mode] operation (Section 3.2).
Splice loss estimation and actual value are inconsistent.
Splice loss estimate is an internal calculated value of the fusion splicer,only
for reference.
Splicer optical components may need to be cleaned.