EasyManua.ls Logo

Snap-On Solus - Viewing Saved Data

Snap-On Solus
87 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
Operations Viewing Data Graphically
Figure 5-24 Sample manually scaling a parameter
3. Use the down d arrow to move the maximum displayed value into position.
4. Press Y/a to set the selected value.
5. Press Y/a to scale the minimum value.
6. Use the up b arrow to move the maximum displayed value into position.
7. Press Y/a to set the selected value.
8. Press N/X to exit manual scale.
The top and bottom of the graph is now clipped so that only the portion that falls between the
selected minimum and maximum values displays.
z To scale all parameters:
1. Select a parameter.
The parameter menu displays.
2. Select Scale All.
The minimum and maximum graph values are reset for all of the parameters.
z To set trigger levels:
1. Select a parameter.
The parameter menu displays.
2. Select Set Trigger Levels.
The active trigger level line displays as a solid black line together with an on-screen position
value (Figure 5-25).
Figure 5-25
Sample active trigger line
NOTE:
i Both an upper and lower trigger level must be set.
3. Adjust the upper trigger level using the up b and down d arrows to place the trigger level line
where you want it on the PID graph.

Table of Contents

Related product manuals