9 Technical data
100 SOREDEX
206718 rev 9
Coefficient of variation of DAP < 5%
Automatic Dose Rate Control
reproducibility
> 95%
X-ray gen
erator
User interface
Program and technique factors
selection, exposure control
Touch screen panel, optional remote exposure switch
Patient positioning Positioning panel, integrated
Connection cable
(CRANEX 3D - PC)
CAT6 UTP Ethernet cable
Panoramic programs & Technical factors & magnification:
Standard Adult Panoramic 57-90 kV/ 3.2-16 mA/2.5-16.4 s 30%
Child Panoramic 57-90 kV/ 3.2-16 mA/2.5-14.4 s 30%
TMJ PA 57-90 kV/ 3.2-16 mA/5.6-10.6 s 23%
TMJ Lateral 57-90 kV/ 3.2-16 mA/5.3-10.6 s 55%
Maxillary Sinus 57-90 kV/ 3.2-16 mA/12.5 s 30%
Bitewing (BW) 57-90 kV/ 3.2-16 mA/11.9 s 30%
Panoramic QC 57-90 kV/ 3.2-12.5 mA/16.4 s 30%
Exposure Control
Exposure Control Pre-programmed icons for all programs
Automatic Spine Compensation
Cephalometric programs & Technical factors:
Reduced width lateral view 85-90 kV / 8-12.5 mA / 6.9-14 s
Full width lateral 85-90 kV / 8-12.5 mA / 10-20 s
PA/AP, facial and oblique views 80-90 kV / 8-12.5 mA / 10-20 s
Carpus program (optional)
(Not in USA and Canada)
60-90 kV / 3.2-12.5 mA / 8-20 s
Exposure Control Automatic Exposure Setting (AES) based on patient
head size
Magnification factor 1.15 (15%)