Vibration measurement E:13
Making a vibration assignment
A proper measuring point for FFT with symptoms
or EVAM has to be made in Condmaster using the
Condition Manager.
In Leonova, you can set up an vibration assign-
ment, returning a set of condition parameters
and single spectrum measured in accordance with
the parameters you input under ‘Measuring point
data’.
You can rename the file with a descriptive name.
Markthevibrationdefaultfile(1)withtheLEFT/
RIGHT arrow keys. Press SHIFT+F2 and select
‘Rename’(2).Inputanameviathekeyboard
window and save with ENTER. The new name will
show up in the measurement window. The default
file(1)willhavethenewnameafterrestartingthe
instrument.
1
2
PressENTERtoopenthevibrationdefaultfile(1).
Press SHIFT+F3 to open ‘Measuring point data’.
Before starting vibration measurements, make sure
that the vibration transducer you are using with your
Leonova is properly selected.
Default transducers for the vibration techniques are
set up via the transducer register. Default transducer
is the active transducer when ‘Portable’ transducer
is selected under ‘Measuring point data’. To select a
default transducer, see Chapter A page 15.
The transducer data become editable after you
switch‘Transducer’(3)from‘Portable’to‘Remote’.
Provided you know the transducer’s frequency range
and its upper and lower bias range, you can input
the data here.
The nominal sensitivity of a vibration transducer
SLD144 is 10.0 mV/m/s
2
. The actual sensitivity of the
individual transducer is written on its calibration
card. This data should always be input in the trans-
ducer register. When several transducers are in use,
they should be marked to assure that the readings
are calibrated.
3