General instrument functions A:19
Create measurement files
The file ‘Create default settings’ (1) under ‘Set-
tings’ is very important. It creates measurement
files for all measuring techniques and places
them under the measuring technique windows.
You cannot use Leonova as a stand-alone meas-
uring instrument without these files.
The installation is simple: Press MENU and open
‘Settings’. Select ‘Create default settings’ and
press F1 (OK).
The example shows the measurement files cre-
ated in the vibration technique window: a file for
vibration ‘1000Hz/1600 lines’ and a file each for
vibration measurement according to ISO 10816
and ISO 2372.
1
Function and use
Leonova has a number of ‘platform’ functions which
are always available. Other functions are user
selectable and can be bought in retrospect when
needed.
‘Function and use’ under ‘MENU/ Settings’ shows
a list of all functions (2), each followed by an icon
showing its status:
Available functions.
Not available in this instrument.
Loaded measuring rounds are shown at the bottom
of the list.
The loading of new functions is described under
’Leonova service program’.
2