Vibration measurement E:19
Measurement results
Press the M/S key to measure or F1 ‘Measure
all’ for a measurement in a round. If the rpm
assignment is set up with ‘Pseudo Tac’ in
Condmaster, this will automatically be acti-
vatedandindicatedtogetherwiththerpm(1).
The ‘Pseudo Tac’ function is described in the
CondmasterUserGuide.
Thelivespectrumdisplay(2)inthe‘Meas-
uring’ window can be changed to show time
signal by pressing Shift+F1.
After measuring vibration, Leonova will
display three main condition parameters:
DISP,VELandACCas(RMS),(Peak)or(Peak
toPeak).Unitsforthedisplayedparameters
are selected under ‘Time signal unit’ when
defining the assignment under ‘Measuring
pointdata’(SHIFT+F3).
Thestatusdots(3)aregreybecausethereare
no evaluation criteria for EVAM assignments
set up in Leonova.
Press F1 to see the full list of condition param-
eters(4)inthemeasuringresultwindow.
To delete a measuring result, press F1
(previous)orF2(next)toselectresultandthen
press F3 to delete.