E:24 Vibration measurement
Measurement results
Press the M/S key to measure or F1 ‘Measure
all’ for a measurement in a round. If the rpm
assignment is set up with ‘Pseudo Tach’ in
Condmaster, this will automatically be activated
andindicatedtogetherwiththerpm(1).The
‘Pseudo Tach’ function is described in the
CondmasterUserGuide.
After measuring, Leonova will display the
result in HDrp (HDrealpeak),ascalarvalue
expressed in decibels. Representing the
true highest peak found in the enveloped
signal, HDrp is the primary value to use for
determining the severity of a bearing or gear
damage. It is also used for triggering alarms in
Condmaster.
In the example two assignments are setup with
differentfitersettings,500-10000 Hz(filter3)
and5000-40 000 Hz(filter4).Bothassignments
are measured parallel when pressing ‘Measure
all’(2).
Thestatusdots(3)aregreybecausethereare
no evaluation criteria for the HD ENV assign-
ments set up in Leonova.
Todeleteameasuringresult,pressF1(Meas-
uringresult),selectresultandthenpressF3to
delete. Selected result is marked with a blue
vertical line.
Press F2 to see the HDrp trend in full scale
where you can re-scale the axis, set comments
etc.(seepartBofthismanual).