E
E:26 Vibration measurement
Motor current analysis
BNC adapter
93487
CAB 97IS
Current clamp
(Flukei1000s)
2
3
1
4
Motor current analysis can be performed with the
EVAM vibration analysis technique using a current
clamp(e.g.Flukei1000s)orapermanentlyinstalled
current tranformer connected to the VIB input.
This feature is used for troubleshooting, primarily
onelectricalmotorswithatleast70%workload.
NOTE: It is important to follow safety precau-
tions and that the measurement is performed
by authorized personnel.
In order to use a current clamp, a user defined
transducer is to be set up. Fluke i1000s is the type
of current clamp used in this example.
Toregisterauserdefinedvibration,pressMENU
andopenSettings’(SHIFT+F3).Select‘Vibra-
tiontransducers’andpressF1(OK)toopenthe
transducerregister(1).
Select‘New’(2)withtheF1key,theninputthe
following data. Press F1 to save the settings and
closewiththeBACKkey.
Name: A descriptive name. It will
be shown on the list of
transducers.
Type: Select‘Userdefined’
Userdefinedname:Isthenameshowninthe
measurement window
Sensitivity: The transducer’s nominal
sensitivity in mV/displayed
unit
Max. frequency: The transducer’s upper
frequency range.
IEPE type: Select ‘No’
Informationabout Sensitivity’ (depends on
whichrangeismeasured)and‘Max.frequencyis
marked on the clamp and/or found in the technical
specificationbelongingtothedevice).
Press F4 ‘Default transducers’, select ‘Vibration’
and press F1. Open the register with ENTER and
select the current clamp to be used with UP/
DOWN. Press F1 to save the settings and close
withtheBACKkey.