A:28 General instrument functions
A
Technical specifications
Instrument, general
Housing: ABS/PC/ TPE
Protection class: IP65
Dimensions: 306 x 108 x 71 mm
(12.4” x 4.3” x 2.8”)
Weight: 860 g (30.3 oz.)
Keypad: sealed, snap action
Display: TFT colour, 240x320 pixels,
3.5 inch, adjustable backlight
Main processor: 400 MHz ARM
®
Memory: 256 MB RAM, 512 MB Flash
SD memory card: 1GB
Operating system: Microsoft Windows
®
CE
DSP processor: 375 MHz floating point
Dynamic range: up to 120dB,
24bit A/D converter
Communication: USB 2.0
Input channels: 1 x VIB, 1 x SPM,
1 x rpm, 1 x temperature
Input/output, headset: 3.5 mm stereo plug
Input/output,
communication: Mini USB
Power supply: rechargeable Lithium-Ion
battery pack 5200 mAh or
power adapter
Battery power: for min. 18 hours normal use
(at 20°C)
Charge time: < 4 hours
Operating temperature: -20 to 55 °C (-4 to 122 °F)
non condensing
Charging temperature: 0 to 45 °C (32 to 113 °F)
General features: language selection, customized
view, voice recording, battery
status idication, transducer line
test, metric or imperial units
Meas. point
identification: NFC transponder for
communication with CondID
TM
tags, read distance max. 50 mm
(2 inch)
SPM input
Shock pulse channels: 1
Measuring techniqes: SPM HD, dBm/dBc, LR/HR and
LR/HR HD
Transducer line test: TLQ test
Transducer types: SPM 40000, 42000, 44000,
probe transducer and quick con-
nector transducer
VIB input
Vibration channels: 1
Frequency range: 0 (DC) to 20 kHz
Resolution: max. 12 800 lines
Transducer line test: bias voltage check
Transducer types: SPM SLD or IEPE* (ICP) type
with voltage output < 24 Vpp, or
DuoTech accelerometer. Trans-
ducer supply of 2.5 mA can be
set On/Off
User defined transd.: e.g. current clamp, pressure and
flow sensors with voltage output
Measuring techniqes: ISO 2372, ISO 10816, FFT with
symptoms, EVAM vibration
analysis, HD Enveloping,
balancing
RPM input
Channels: 1, RPM or temperature
Measuring range: 1 to 150 000 pulses/min.
Resolution: 1 pulse
Accuracy: ± (1 pulse + 0.1% of reading)
Transducer type: SPM TTP10, TTL-pulses,
Key phasor
®
, proximity switch
NPN/PNP
Output: TTL signal for controlling
stroboscope and 12 VDC supply
Temperature measurement
Temp. range (TTP10): -20 to 300 °C (-4 to 572 °F)
Accuracy (TTP10): ± 2.5 °C
Transducer type: SPM TTP10
Stethoscope
Settings: Filter, volume and gain
Transducer types: Shock pulse and vibration
transducers

Table of Contents

Related product manuals