General measurement functions B:9
B
‘Recording’ on the measurement menu (1) is a func-
tion for taking a stated number of readings at stated
intervals, or measure for a stated number of minutes.
Default files and single measurement user files can
be used when recording a single quantity, e. g. shock
pulses or a certain type of vibration.
However, the ‘Recording’ feature is consecutive
measurement of different quantities, using up to
three different transducers simultaneously con-
nected to Leonova:
a shock pulse transducer on the SPM input
one vibration transducer on the vib input
a tachometer or temperature probe on the ta-
chometer/temperature input
To set up a consecutive recording of a shock pulse
measurement, a vibration or rpm measurement and
possibly temperature (or any combination of these),
one needs a measuring point where all wanted
measuring techniques are active. This measuring
point is either downloaded from Condmaster, or it
is a multi-measurement user file (see previous page).
Define the number of measurements or minutes (2)
and the time interval (3) between measure ments (0
minutes = as fast as possible). Use the F1 key (NEW,
4) to select measuring techniques from the list (5)
and put them into the measuring sequence (6). A
selected technique can be replaced by another
with EDIT (7) or be deleted with (8). Connect the
transducer(s) and press the MEASURE/SAVE (M/S)
key to start.
The results can be seen on the graphics display and
can be uploaded to Condmaster.
For SPM and vibration measurements with variable
speed, select ‘Speed’ as a technique and do not
input an rpm under ‘Measuring point data’.