Vibration severity ISO 2372
Measurement
quantities: velocity, RMS value in mm/s
over 10 to 1000 Hz
Evaluation table
selection: menu guided, ISO 2372
Transducer line test: bias voltage test
Vibration severity ISO 10816 with spectrum
Measurement
quantity: velocity, acceleration, and
displacement according
to individual parts in the
standard
Spectrum, linear, 1600 lines, Hanning
window.
Quick mode: yes (can be set ON/OFF)
Transducer line test: bias voltage test
HD ENV
Spectrum lines, frequency and envelope filter limits are
regulated by what is included in the vibration package.
Frequency limit, upper: stated in orders
Envelope high pass filters: 100, 200, 500, 1000, 2000,
5000, 10 000 Hz
Envelope band pass filters: Filter 1 = 5 -100 Hz
Filter 2 = 50 -1000 Hz
Filter 3 = 500 -10 000 Hz
Filter4=5000-40000Hz*
Averages: Time synch
Spectrum lines: 400, 800, 1600, 3200, 6400,
12800
Frequency units: Hz, CPM, orders
Saving options: full spectrum, time signal and
FFT, condition parameters
Vibration analysis, EVAM/FFT with symptoms
Frequency limit,
lower: 0 (DC) to 200 Hz
Frequency limit,
upper: 8 to 20 000Hz (EVAM)
8 to 10 000 Hz (FFT)
order tracking
Envelope high
pass filters: 100, 200, 500, 1000, 2000,
5000, 10 000 Hz
Envelope band
pass filters: 5-100, 50-1000, 500-10000,
5000-40000 Hz
Measurement
windows: Rectangle, Hanning,
Hamming, Flat Top
Averages: time synchronous, FFT linear,
FFT peak-hold
Spectrum lines, EVAM: 400, 800, 1600, 3200, 6400,
12800
Spectrum lines, FFT: 400, 800, 1600, 3200, 6400
Saving options: time signal (FFT calculated),
full spectrum, time signal and
FFT, condition parameters
Spectrum types
displayed: linear, power, PSD
Zoom, EVAM: true FFT zoom, visual zoom
Zoom, FFT: visual zoom
Quick mode: yes (can be set ON/OFF)
Frequency units: Hz, CPM, orders
Trigger selection: RPM trigger, post trigger
Transducer line test: bias voltage test
*IntegralElectronicPiezoElectric
Specifications are subject to change without notice.