# Error message Explanation Action
80 Error
The mover head did not
move correctly.
- Check the air supply.
- Check for any
obstructions.
Something has prevented the
specimen mover head tried
from moving.
Check the air supply.
Make sure that nothing
obstructs the movement
of the specimen mover
head.
99 Error
Disc motor:
Movement is blocked.
Motor and/or movement is
blocked.
Make sure that nothing
obstructs the movement.
100 Error
Disc motor:
Error: General.
Error code: 0
Try to restart the
machine
Non-specific error reported by
movement unit.
Re-start.
If the error remains,
contact Struers Service.
Make a note of the error
code displayed.
103 Error
The required air
pressure is not
obtained.
- Check the air supply.
Make sure that the air
pressure is min. 6 bar.
Re-start.
If the error remains,
contact Struers Service.
118 Error
Uploading of new
consumables failed! Try
uploading the
consumables again.
Consumables has been
reset to default.
An error occurred during
upload of the consumables
table from the PC to the
machine.
Check the connection
between the PC and
machine.
161 Error
Disc motor:
Frequency inverter
temperature alarm.
Reason code: 0x0
The load on the frequency
inverter which drives the disc
motor has caused a
temperature alarm.
Reduce the load.
52 AbraPol-30
8 Troubleshooting - AbraPol-30