# Error message Explanation Action
29 Fatal error
Emergency stop mode
active, but monitoring
switch released.
The emergency stop button
has been activated, but the
internal monitoring switch is
not active.
This error can appear if
the emergency stop
button is released very
slowly (i.e. more than
several seconds).
Re-start.
If the error remains,
contact Struers Service.
44 Fatal error
Disc motor:
Communication error.
Call a Service
Technician.
Communication with the disc
motor frequency inverter has
been lost.
Re-start.
If the error remains,
contact Struers Service.
61 Fatal error
Machine failed during
Power On Self Testing.
Try restarting the
machine.
Contact Struers
technical support if the
problem persists.
Reason: #__ - Unknown
error
Re-start.
If the error remains,
contact Struers Service.
Make a note of the
reason code.
62 Fatal error
The cover open signal is
present while the lock is
locked.
Call a Service
Technician.
Make sure that the cover
is completely closed.
Re-start.
If the error remains,
contact Struers Service.
77 Fatal error
Emergency stop
released, but monitoring
switch still on.
Call a Service
Technician.
Re-start.
If the error remains,
contact Struers Service.