Articulating Head Piece Adjustment
Required Tools:
#2 Phillips Screwdriver 7/16” Socket and Ratchet Standard Screwdriver
Adjustment Procedure:
1. If the neck section is binding, the (patient) left handle needs adjustment. If the head section is binding,
the (patient) right handle needs adjustment. If you are adjusting the neck section handle, the 4 Phillips
head screws (A) holding the neck section cover (B) must be removed.
2. If the head or neck section is not returning to the full lock position, use a 7/16 socket and ratchet to loosen
the pivot nut (C).
3. If the torque block (E) is binding, use a 7/16 socket and ratchet to loosen the pivot nut (D).
4. If the section is still binding, check the torque block (E) for excess pressure against the gear (F). Using
a 7/16 socket and ratchet, loosen the two nuts (G) holding the gear adjustment block. Tighten the nuts
and verify the handle is releasing properly.
CAUTION
The dual articulating headpiece is designed to provide precision surgical positioning. Be sure to treat it with
care. The unit should be routinely checked to ensure optimal performance. In the event of any impact or
overload of the headpiece, be sure the unit is working properly and supports the intended load. Verify the
adjustment gears lock and release properly and, if necessary, perform the adjustment procedure.
A
A
B
C
C
D
F
F
G
G
E
E