www.stryker.com 2030-809-002 REV D 15
Bed Serial Number:
Completed by: _______________________________________ Date: _________________
Preventative Maintenance
Beds require an effective maintenance program, we recommend checking these items annually. Use this sheet for
your records. Keep on file.
CHECKLIST
_____ All fasteners secure.
_____ Engage brake pedal and push on the frame to ensure all casters lock securely.
_____ Inspect the brake assembly (Brake Cam, Brake Plate Body, Brake Ratchet Spring and Brake Bar) for
degradation or signs of wear at the foot end and head end of the bed. Ensure brake assembly components
are functioning properly.
_____ Locking steer caster engages and disengages properly - optional equipment.
_____ Siderails move, latch and stow properly.
_____ All functions on siderails working properly (including LED’s).
_____ Head End Control Panel working properly (including LED) - optional equipment.
_____ Confirm battery powered functionality - optional equipment.
_____ Manual CPR release working properly.
_____ Optional foot prop intact and working properly.
_____ I.V. pole working properly.
_____ Foley bag hooks intact.
_____ Chart rack intact and working properly.
_____ CPR board not cracked or damaged and stores properly.
_____ No cracks or splits in head and footboards.
_____ All functions on footboard working properly (including LEDs).
_____ No rips or cracks in mattress cover.
_____ Scale and Bed Exit system calibrated properly - optional equipment.
_____ Motion interrupt switches working properly.
_____ Power cord not frayed.
_____ No cables worn or pinched.
_____ All electrical connections tight.
_____ All grounds secure to the frame.
_____ Ground impedance not more than 100 mΩ.
_____ Current leakage not more than 300 µA.
_____ Apply grease to the bed grease points including the fowler clutch and brake cam.
_____ Ensure ground chains are clean, intact, and have at least two links touching the floor.
_____ “Brake” LED on the footboard (iBED Awareness option only) blinks when brakes are not engaged.
_____ Siderail switches working properly (iBED Awareness option).
_____ Bed Status Light Bar LEDs working properly (iBED Awareness option).
_____ Bed Status Side Indicator LEDs working properly (iBED Awareness option).
_____ Inspect footboard control labeling for signs of degradation (iBED Awareness option).
_____ Optional accessories intact and working properly.
_____ Check labels, as specified in the Operations and Maintenance manuals to ensure legibility, proper adherence
and integrity.
_____ Fowler functioning properly.