72 2030-809-002 REV D www.stryker.com
FOOTBOARD LID REMOVAL
Tools Required:
• #2 Phillips Screwdriver
Procedure:
1. Using a #2 Phillips screwdriver, remove the screws (A & B) holding the door and hinge assembly to the
footboard.
2. If replacing the Lid only, use a Phillips screwdriver to remove the screws holding the hinge to the door.
3. Reverse the above steps to attach the replacement door and/or hinge.
Note
Screw (B) is a machine screw and must be reinstalled in the proper hole.
NURSE CALL 9V BATTERY REPLACEMENT
If the Power LED on the footboard is flashing, the 9V Nurse Call battery located under the actuator box needs to be
replaced. No tools are required to replace the battery. Unplug the power cord from the wall socket and replace the bat-
tery. After replacing the battery, verify the Power LED is no longer flashing. Dispose of the old battery in accordance
with local regulations.