57 EN
Voice Commands Menu
Awaken
Device
Select
Control Select Description
Sidne™
Arthro
Pump
Pressure
Ten Up Increases the pump pressure by 10 mm Hg.
Ten Decreases the pump pressure by 10 mm
Up Increases the pump pressure by 5 mm Hg.
Down Decreases the pump pressure by 5 mm Hg.
Activate Starts the pump.
Wash
Off Turns the wash mode off.
On
Turns the wash mode on and activates the
pump.
Flow
Up Increases the flow by 0.25 L/min.
Down Decreases the flow by 0.25 L/min.
Joint
Shoulder Selects shoulder joint type.
Knee Selects knee joint type.
Small Selects small joint type.
Stop Stops the pump.