Electromagnetic Compatibility Tables
75 EN
Guidance and Manufacturer's Declaration: Electromagnetic
Immunity
Sidne™ is intended for use in the electromagnetic environment
specified below. The customer or the user of Sidne™ should ensure
that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment:
Guidance
Electrostatic
Discharge (ESD)
IEC61000-4-2
±6kV
contact
±8kV air
±2,4,6kV
contact
±2,4,8kV air
Floors should be
wood, concrete,
or ceramic tile.
If floors are
covered with
synthetic
material, the
relative
humidity should
be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2kV for
power
supply lines
±1kV for
input/output
lines
±0.5,1,2kV
for power
supply lines
±0.25,0.5,1k
V for input/
output lines
Mains power
quality should
be that of a
typical
commercial or
hospital
environment.
Surge
IEC61000-4-5
±1kV
differential
mode
±2kV
common
mode
±0.5, 1kV
differential
mode
±0.5,1,2kV
common
mode
Mains power
quality should
be that of a
typical
commercial or
hospital
environment.