Electromagnetic Compatibility Tables
77 EN
Guidance and Manufacturer's Declaration: Electromagnetic
Immunity
Sidne™ is intended for use in the electromagnetic environment
specified below. The customer orthe user of Sidne™ should ensure
that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment:
Guidance
Conducted
RF
IEC
61000-4-6
Radiated
RF
IEC
61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80MHz to
2.5 GHz
3 V
3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Sidne™
system, including its
cables, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
Separation Distance
80 MHz to
800 MHz
800 MHz to
2.5 GHz
d1.2P=
d1.2P=
d2.3P=