#1,U973,N1234,W1.01.1,Q0.01,M1,R1,F2:1,F3:2,F1:3,F2:4,F3:5,F1:6,J2:1,J3:2,J1:3,J3:4,J3:5,J1:6,f1
,C1:1,C0:2,C2:3,C1:4,C0:5,C2:6,B0:1,B3:2,B15:3,b0,d1s,D10s,K5,L0,Y3,y0,XT0,XL100,XQ0,Xq0,XC
115:1,XC115:2,XC115:3,Xl115:1,Xl115:2,Xl115:3,XA0,XD-1:1,XD-1:2,XD-1:3,XD-1:4,XD-1:5,XD-
1:6,S0,T1,e480,c1:1,c1:2,c1:3,h0:1,h0:2,h0:3,x3:1,x3:2,x5:3,m0,s0,l100,O10,o0,t0;
means that:
• SV 973 is investigated (U973);
• its number is 1234 (N1234);
• software version number is 1.01.1 (W1.01.1);
• calibration factor is equal to 0.01 dB (Q0.01);
• LEVEL METER is selected as the measurement function (M1);
• range is LOW (R1);
• A filter is selected in profile 1, SLM function (F2:1);
• C filter is selected in profile 2, SLM function (F3:2);
• Z filter is chosen in profile 3, SLM function (F1:3);
• A filter is selected in profile 1, DOSE function (F2:4);
• C filter is selected in profile 2, DOSE function (F3:5);
• Z filter is chosen in profile 3, DOSE function (F1:6);
• A Peak filter is selected in profile 1, left channel, SLM function (J2:1);
• C Peak filter is selected in profile 2, left channel, SLM function (J3:2);
• Z Peak filter is selected in profile 3, left channel, SLM function (J1:3);
• C Peak filter is selected in profile 1, both channels, DOSE function (J3:4);
• C Peak filter is selected in profile 2, both channels, DOSE function (J3:5);
• Z Peak filter is selected in profile 3, both channels, DOSE function (J1:6);
• Z filter is selected for 1/1 OCTAVE or 1/3 OCTAVE analysis (f1)
• FAST detector is selected in profile 1, SLM function (C1:1);
• IMPULSE detector is chosen in profile 2, SLM function (C0:2);
• SLOW detector is selected in profile 3, SLM function (C2:3);
• FAST detector is selected in profile 1, DOSE function (C1:4);
• IMPULSE detector is chosen in profile 2, DOSE function (C0:5);
• SLOW detector is selected in profile 3, DOSE function (C2:6);
• logger’s buffer is not filled by the results from profile 1 (B0:1);
• Lpeak and Lmax values are stored in the files of the logger from profile 2 (B3:2);
• Lpeak, Lmax, Lmin and Leq values are stored in the files of the logger from profile 3 (B15:3);
• results of 1/1 OCTAVE or 1/3 OCTAVE analysis are not stored in the files of the logger (b0);
• results are stored in a logger’s file every 1 second (d1s);
• integration period is equal to 10 seconds (D10s);
• the measurement has to be repeated 5 times (K5);
• linear detector is selected to the Leq calculations (L0);
• delay of the start of the measurements is equal to 3 seconds (Y3);
• synchronization the start of measurement with RTC is switched off (y0);
• logger triggering mode is switched off (XT0);
• logger triggering level is set to 100 dB (XL100);
• number of the records before the triggering saved in a file of the logger is equal to 0 (XQ0);
• number of records registered, after the moment in which the measured signal does not fulfil
any longer the condition of the triggering, is equal to 0 (Xq0);
• threshold level for PTC calculation in profile 1, is set to 115 dB (XC115:1);
• threshold level for PTC calculation in profile 2, is set to 115 dB (XC115:2);