XLi Time & Frequency System 113
XLi-man, Issue 8, 6/17/2008, Rev. H
2
5
1
SSSSSSSSSSSS SSSSSSSSSS SSSSSSS S SSS S SSSSS S
FPGA 230-01510-38v01<CR><LF>
To obtain the Parallel BCD Output option board BCD Time Mode information, send the following string to
the Serial/Network port:
F78<SP>B<N><SEP>TM<CR><LF>
where:
F = ASCII character
78 = function number.
<SP> = ASCII space character one or more.
B = ASCII character to denote Option Bay number follows
<N> = Option Bay Number, 1 through 10.
TM = ASCII character for TM selection request
<LT> = line terminator; for output strings (a carriage return and line feed ) or for input strings (a
carriage return only).
Sample request: F78 B2 TM<CR><LF>
Response:F78 B2 TM LOCAL <CR><LF>
To change the Parallel BCD Output option card BCD Time Mode selection, send the following send the
following string to the Serial/Network port:
F78<SP>B<N><SP>TM<SP><OUT><LT>
where:
F = ASCII character F.
78 = function number.
<SP> = ASCII space character.
B = ASCII character to denote Option Bay number follows
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com