6-1-2-6-4. Pulse displacement sensor
Content Command No. Setting item
Count number (higher 2 digits) 1600
Higher 2 digits (1–15000000)
o Set the higher digits followed by the lower digits.
Count number (lower 6 digits) 1601 Lower 6 digits (1–15000000)
Display value (equivalent input) 1602 00001–99999
Display value (actual load) 1603 00001–99999
Zero position 1604 ±99999
Moving average number (MOVING AVG. NUM.) 1605 0–1: Disabled, 2–2048: Moving average number
Output phase (AB, A) 1606 0: AB phase, 1: A phase
6-1-2-6-5. Voltage displacement sensor
Content Command No. Setting item
Rated output 1610 100–5200 (0.100–5.200V)
Display value (equivalent input) 1612 00001–99999
Display value (actual load) 1613 00001–99999
Zero position 1614 ±99999
Moving average number (MOVING AVG. NUM.) 1615 0–1: Disabled, 2–2048: Moving average number
Low-pass filter 1616 0: 10Hz, 1: 30Hz, 2: 100Hz, 3: 300Hz
6-1-2-6-6. TEDS
Content Command No. Setting item
Serial number 6001 0–9999999
Maximum rated capacity 6002 Decimal point position + 5 digits
Maximum rated output 6003 Decimal point position + 5 digits (mV/V)
Sensor impedance 6004 Decimal point position +5 digits (Ω)
Max. Excitation level 6005 Five digits (V)
Calibration date (CAL. DATE) 6006 Year/month/day (yyyy/mm/dd)
Model number 6007