5 - 4 Te-Stack Operating Manual, 391862, en, V2.3
5 - Putting into Operation
Reinstalling the Te-Stack
5.2.4 Settings and Checks after Reinstallation
5.2.4.1 Te-MO Version
1 Perform the setup Mechanical Adjustment of Te-Stack positions on the Te-
MO panel.
Refer to the “Instrument Software Manual”.
2 Perform the setup Teach Te-Stack positions in the Te-Stack panel.
Refer to the “Instrument Software Manual”.
5.2.4.2 Freedom EVO Version
Teach the RoMa position.
Refer to the “Freedom EVOware Software Manual”.
5.2.4.3 All Versions
Checking the Carrier Positions
Refer to chapter 7.5.4 “Teaching the Stack Position”, 7-6.
Requested Tests and Settings
Refer to chapter7.4 “Performing the Requested Tests and Settings”, 7-4
and to the “Instrument Software Manual”.
Perform the Range Test procedure
Perform the Mechanical Burn In / Validation procedure
Perform the Te-Stack Functional Test procedure
Perform the Transfer Station Functional Test procedure
Perform the System Test procedure