7 - Preventive Maintenance
Adjustments and Replacements
Te-Stack Operating Manual, 391862, en, V2.3 7 - 5
7.5 Adjustments and Replacements
7.5.1 Safety Instructions
WARNING
Chemical, biological and radioactive hazards can be associated with the
substances used or the samples processed with the Te-Stack.
The same applies to waste disposal.
Always be aware of possible hazards associated with these substances.
Use appropriate protective clothing, safety goggles and gloves.
WARNING
Potential mechanical hazard
Mount and close all covers before switching on the instrument.
Keep out of the moving areas during the time the instrument is switched on.
All the safety measures of Freedom EVO and Te-MO apply also for Te-Stack.
Attention
Discharge static electricity from your body with a bracelet. Keep it on during all
maintenance tasks, including setup & service procedures, i.e. while working on
the instrument and its equipment to protect yourself and sensitive equipment
installed.
7.5.2 Removing the Acrylic Glass Covers
7.5.2.1 Te-MO Version
Note: A special front cover is mounted on the Te-MO.